<div id="popup_content" title="LRS 7126" last_modified="2011-07-24T17:07:34+00:00">

	<p><img src="../images/me/cosmic_rays/air_shower_icon.gif" alt="air shower" class="img_title_icon"/> <span class="title">Level Translator</span></p>

	<p><img src="../images/me/cosmic_rays/LRS_7126.jpg" alt="LeCroy 7126 Level Translator" class="float_left"/>
	The LeCroy 7126 Level Translator translates one voltage standard to another.
	Its inputs and outputs can be NIM, TTL, or ECL.</p>

	<table cellpadding="5" cellspacing="0" border="1">
		<tr align="center">
			<th></th>
			<th>Logical 0</th>
			<th>Logical 1</th>
		</tr>
		<tr align="center">
			<th>ECL</th>
			<td>-0.90 V</td>
			<td>-1.75 V</td>
		</tr>
		<tr align="center">
			<th>NIM</th>
			<td>0 V</td>
			<td>-0.8 V</td>
		</tr>
		<tr align="center">
			<th>TTL</th>
			<td>0 V</td>
			<td>5 V</td>
		</tr>
	</table>

	<p>Delay: NIM, TTL, or ECL -&gt; NIM or ECL output = 5ns<br/>
	Delay: NIM, TTL, or ECL -&gt; TTL output = 10ns</p>